Dr. Födisch Biogas-Monitor

The biogas monitor can monitor up to 4
gas components continuouslyDr. Födisch Biogas-Monitor

Bio Gas Monitor BGM 06

  • continuous measurement of max. 4 gas components
  • measuring principle based on dilution
  • for determination of the energy content (CH4 measurement) 
  • for protection of engines (H2S measurement)
Fields of application 

Operating a modern biogas plant is hardly possible without a continuous control of the gas composition. Knowing the gas composition helps the operator to increase the efficiency of the plant and to counteract promptly wrong developments. In addition more and more manufacturers of combined heat and power units demand for the compliance with pollutant limit values - especially for the aggressive H2S - in order to keep warranty.

The biogas monitor can monitor up to 4 gas components continuously. The basic version includes the measurement of CH4 (methane) and CO2 (carbon dioxide) by means of an infrared cell as well as O2 (oxygen) and H2S (hydrogen sulphide) by means of an electrochemical cell.

 Technical data

Case: 3-part wall case; dimensions: 610 x 750 x 480 mm (W x H x D),  weight ca. 25 kg

Ambient
temperature: +5 … +40 °C
Supply Voltage: 230 VAC/50-60 Hz
Probe

Analyzer Gas sampling via flame Barrier

principle: mix of sample gas with inert gas in proportion of 1:50

Measuring Infrared photometer (CH4, CO2)
principle Electro-chemical cell (O2, H2S)


Measuring Ranges:
CO2   0 - 100 Vol%
CH4 0 - 100 Vol%
H2S 0 - 10.0000 ppm
02 0 - 25 Vol%

Display: LC-Display, 240 x 128 pixel, back-lighted
Keyboard: Keypad
Operation: Menu-driven
Detection Limits < 2 % of the respective measuring range
Zero Point Correction: Automatically
Baro Correction: Internally
Response time: T90 < 180 s (depending on plant and chosen component)
Drift: < 5% in maintenance interval
Analogue outputs: 4 ... 20 mA (CO2 , CH4 , H2S and O2)
Digital signals: Failure, maintenance, maintenance requirement, limit values
Interfaces: RS 232, Profibus (optional)
Maintenance: At least half-yearly

Documentation

Dr. Födisch Product information Biogas-Monitor Dr. Födisch Product information Biogas-Monitor

Ankersmid engineers, delivers and performs service on analysis equipment, to complete CEM systems and process systems in associated analysis houses. They are used in industrial process technology, laboratories, air force, stable technology and biogas installations

Ankersmid Process
Samuel Morsestraat 4
7442 DH Nijverdal
Netherlands


(31)-0548-59 59 69
(31)-0548-62 19 72
info@ankersmid.com

© 2024 - Ankersmid Group